Two-dimensional electron cyclotron emission imaging diagnostic for TEXTOR Journal Article uri icon

Overview

abstract

  • A two-dimensional electron cyclotron emission imaging (ECEI) diagnostic has been developed in which broad bandwidth radiation is collected by a vertically aligned mixer array and separated by frequency band. Using a 16 element array, and an 8 band receiver attached to each array element, time-resolved 16×8 images of electron temperature profiles and fluctuations of the TEXTOR plasma are acquired with vertical and horizontal channel spacings of 11 and 8 mm, and spot sizes of 13 and 9 mm, respectively. The system is wideband tunable from 95 to 130 GHz, and the focus location may be shifted horizontally via translation of one of the optical imaging elements. System design and laboratory testing details of the ECE Imaging optics and multichannel wideband electronics are presented, together with TEXTOR plasma data.

publication date

  • October 1, 2004

has restriction

  • closed

Date in CU Experts

  • February 19, 2015 2:55 AM

Full Author List

  • Wang J; Domier CW; Xia ZG; Liang Y; Luhmann NC; Park H; Munsat T; Mazzucato E; van de Pol MJ; Classen IGJ

author count

  • 11

Other Profiles

International Standard Serial Number (ISSN)

  • 0034-6748

Electronic International Standard Serial Number (EISSN)

  • 1089-7623

Additional Document Info

start page

  • 3875

end page

  • 3877

volume

  • 75

issue

  • 10